YZ 200 , YZ 300

YZ 200MA , YZ 300MA

صيني ، ترابيزة متحركة ، وحدة تحكم انالوج



YZ - 200 B ( 200 mA ) & YZ - 300 ( 300mA )
MAIN TECHNICAL PARAMETER
Single bed with single bulb
Design according to the electric mechanism of middle-size machine, with complete and reasonable function
High-voltage system is of split structure with high voltage generator of full wave rectification, independent X-ray tube unit, large output power and reliable and stable operation
Power voltage (V) photograph kilovolt (KV) and infinitely variable control
Be equipped with the Monistat for the filament of X-ray tube and space complementary to ensure the stable and exact output
Be equipped with capacity limit protective circuit --- kV mAS interlock limit protective
The timer circuit grades according to R 10 priority coefficient has an exact time control
The photographic flat bed has a vibrating ray-filter
The pillar has only the bottom track with easy installation
General Photography
Filter-ray device photography
ITEMCONTENT                                    INDEX
Demand of power supplyVoltage220 V / 380 V + 10% single Phase 220 V/380 V+ 10%
Freguency50 Hz + 0.5 Hv
Capacity10 KVA
Internal resistance380 V : 2 W ; 220V : 0.6 W 0 2W at 380V;0.6Wat 220 V
PhotographyVoltage50 ~ 90 KV                      50 ~ 100 KV
Current
25 mA       50 mA        75 mA      100 mA    125 mA
Small focus : 25 mA, big focus 50 mA 75 mA 100 mA 125 mA (grading)
Time0.04 ~ 6.3 s 23   .0.04 ~ 6.3 s, 23 grades together
High-voltage transformer unitCapacity             7 KV                        8.4 KVA
Max.Dc output voltage             90 KV                     100 KV
Max.Dc output current            100 mA                  125 mA
X-Ray tubeModelXD4-2.9/100
Focus1.8 mm   4.3 mm/small focus : 1.8 mm big focus: 4.3 mm
(YZ-SCIII) Photographic bedRay - Filter1960 mm x 660 mm x 700 mm
(YZ-SCIII)N28 / Grid density N28
F0 100 cm / Convergence distance fo 100 cm
R8 / Grid ratio r8
Pillar of X-ray source unitMoving in length (along the bottom track)> 1350 mm
Moving in breadth       > 360 mm
Moving up and down720 ~ 1800 mm
Max size cassette for photograph356 x 432 mm (14” x 17”
  
MAX CAPACITY OF PHOTOGRAPY
Focus of X-ray tubePhotograph mAMax photograph KVMax allowable time
Small focus25 mA90 KV100 KV1.0 s
Big Focus50 mA90KV95 KV6.3 s
75 mA90 KV95 KV5 s
100 mA85 KV90 KV3.2 s
125 mA85 KV1.0 s